期刊
FOOD CHEMISTRY
卷 126, 期 2, 页码 756-761出版社
ELSEVIER SCI LTD
DOI: 10.1016/j.foodchem.2010.11.060
关键词
Biogenic amine; Validation; Tuna fish; High-performance liquid chromatography
资金
- Nutrition and Food Science Faculty of Shahid Beheshti Medical Science University of Iran
This study compared and validated two methods of high-performance liquid chromatography: fluorescence detection (HPLC-FLD) and UV-Vis detection. Recoveries greater than 55% at all spiking levels using UV-Vis detection could not be obtained. However, derivatization with pre-column o-phthalaldehyde (OPA) was found to efficiently enhance the method sensitivity. The analytical method was validated based on the following criteria: linearity, sensitivity, accuracy, precision, repeatability, and recovery. For fluorescence detection, excellent linear correlation with R-2 = 0.9977 was observed over the range of 5.0 to 100.0 mg/L. The minimum detection limit was calculated to be 1.5 mg/kg, while the limit of quantification (LOQ) value of the validated method was determined to be 4.5 mg/kg. Good recoveries (RSD% = 1.35) were obtained at all spiking levels ranging between 0 and 200 mg/kg. The proposed method was found to be suitable, selective, and rapid for the determination of histamine. (C) 2010 Elsevier Ltd. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。