4.4 Article

Quantitation in gas chromatography: usual practices and performances of a response factor database

期刊

FLAVOUR AND FRAGRANCE JOURNAL
卷 23, 期 6, 页码 450-459

出版社

WILEY
DOI: 10.1002/ffj.1906

关键词

GC; MS; FID; response factors; quantification; database; internal standardization

向作者/读者索取更多资源

In this study, usual methods commonly applied to report the quantitative composition of flavours, fragrances and essential oils were compared. The MS determinations without response factors exhibited a lack of accuracy and reproducibility. FID provided better results, but the determinations were still unsatisfactory. The best results were obtained by true internal standardization in MS and FID, which is time-consuming for mixtures composed of many constituents. The use of a response factor database was thus proposed and validated by testing the response dependence on GC parameters, using an experimental design. Very low time variations were observed. Under standard GC conditions, the determination of a model mixture with four different instruments exhibited a mean bias of < 3.4%. Copyright (C) 2008 John Wiley & Sons, Ltd.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据