4.3 Review

All-inside meniscal repair devices and techniques

期刊

EXPERT REVIEW OF MEDICAL DEVICES
卷 9, 期 2, 页码 147-157

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1586/ERD.11.74

关键词

all-inside; fourth generation; meniscal repair; third generation

向作者/读者索取更多资源

Techniques and devices for meniscal repair are evolving, including for all-inside meniscal repair. The first-generation all-inside meniscal repair devices were simplistic in design, but their technical difficulties and risks led to the development of second-generation devices. These devices employed the suture anchor concept for repair and confirmed the safety of use through standard anterior arthroscopic portals. Third-generation devices introduced the idea of rigid, bioabsorbable materials; unfortunately, these demonstrated higher failure and complication rates compared with other repair techniques. They were also limited in their ability to adjust compression and tension across the repair. Now, fourth-generation devices have been developed that are flexible, suture-based, and allow for variable compression and retensioning across the tear. Each device has its own specifications and technical nuances. With a comprehensive understanding of the current devices available, the industry and surgeons may continue in the development of safer, more successful, user-friendly and cost-effective all-inside devices.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据