Improved Differential Ion Mobility Separations Using Linked Scans of Carrier Gas Composition and Compensation Field

期刊

出版社

SPRINGER
DOI: 10.1007/s13361-015-1208-9

关键词

Differential ion mobility spectrometry; DIMS; FAIMS; Peak capacity; Carrier gas; Mass spectrometry

向作者/读者索取更多资源

Differential ion mobility spectrometry (DIMS) separates ions based on differences in their mobilities in low and high electric fields. When coupled to mass spectrometric analyses, DIMS has the ability to improve signal-to-background by eliminating isobaric and isomeric compounds for analytes in complex mixtures. DIMS separation power, often measured by resolution and peak capacity, can be improved through increasing the fraction of helium in the nitrogen carrier gas. However, because the mobility of ions is higher in helium, a greater number of ions collide with the DIMS electrodes or housing, yielding losses in signal intensity. To take advantage of the benefits of helium addition on DIMS separations and reduce ion losses, linked scans were developed. In a linked scan the helium content of the carrier gas is reduced as the compensation field is increased. Linked scans were compared with conventional compensation field scans with constant helium content for the protein ubiquitin and a tryptic digest of bovine serum albumin (BSA). Linked scans yield better separation of ubiquitin charge states and enhanced peak capacities for the analysis of BSA compared with compensation field scans with constant helium carrier gas percentages. Linked scans also offer improved signal intensity retention in comparison to compensation field scans with constant helium percentages in the carrier gas.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据