CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT

期刊

EUROPEAN RADIOLOGY
卷 22, 期 2, 页码 295-301

出版社

SPRINGER
DOI: 10.1007/s00330-011-2271-7

关键词

Computed tomography; Multidetector CT; Dose reduction; Image quality; Iterative reconstruction; Lumbar spine

向作者/读者索取更多资源

Objectives To evaluate the impact of Adaptive Iterative Dose Reduction (AIDR) on image quality and radiation dose in phantom and patient studies. Methods A phantom was examined in volumetric mode on a 320-detector CT at different tube currents from 25 to 550 mAs. CT images were reconstructed with AIDR and with Filtered Back Projection (FBP) reconstruction algorithm. Image noise, Contrast-to-Noise Ratio (CNR), Signal-to-Noise Ratio (SNR) and spatial resolution were compared between FBP and AIDR images. AIDR was then tested on 15 CT examinations of the lumbar spine in a prospective study. Again, FBP and AIDR images were compared. Image noise and SNR were analysed using a Wilcoxon signed-rank test. Results In the phantom, spatial resolution assessment showed no significant difference between FBP and AIDR reconstructions. Image noise was lower with AIDR than with FBP images with a mean reduction of 40%. CNR and SNR were also improved with AIDR. In patients, quantitative and subjective evaluation showed that image noise was significantly lower with AIDR than with FBP. SNR was also greater with AIDR than with FBP. Conclusion Compared to traditional FBP reconstruction techniques, AIDR significantly improves image quality and has the potential to decrease radiation dose. Key Points This study showed that Adaptive Iterative Dose Reduction (AIDR) reduces image noise. In a phantom image noise was reduced without altering spatial resolution. In patients AIDR reduced the image noise in lumbar spine CT. AIDR can potentially reduce the dose for lumbar spine CT by 52%.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据