Volumetric measurement of pulmonary nodules at low-dose chest CT: effect of reconstruction setting on measurement variability

期刊

EUROPEAN RADIOLOGY
卷 20, 期 5, 页码 1180-1187

出版社

SPRINGER
DOI: 10.1007/s00330-009-1634-9

关键词

Multi-detector computed tomography; Pulmonary nodules; Volume measurement; Section thickness; Reconstruction kernel

向作者/读者索取更多资源

To assess volumetric measurement variability in pulmonary nodules detected at low-dose chest CT with three reconstruction settings. The volume of 200 solid pulmonary nodules was measured three times using commercially available semi-automated software of low-dose chest CT data-sets reconstructed with 1 mm section thickness and a soft kernel (A), 2 mm and a soft kernel (B), and 2 mm and a sharp kernel (C), respectively. Repeatability coefficients of the three measurements within each setting were calculated by the Bland and Altman method. A three-level model was applied to test the impact of reconstruction setting on the measured volume. The repeatability coefficients were 8.9, 22.5 and 37.5% for settings A, B and C. Three-level analysis showed that settings A and C yielded a 1.29 times higher estimate of nodule volume compared with setting B (P = 0.03). The significant interaction among setting, nodule location and morphology demonstrated that the effect of the reconstruction setting was different for different types of nodules. Low-dose CT reconstructed with 1 mm section thickness and a soft kernel provided the most repeatable volume measurement. A wide, nodule-type-dependent range of agreement between volume measurements with different reconstruction settings suggests strict consistency is required for serial CT studies.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据