4.7 Article

Covalent immobilization of glucose oxidase on the film prepared by electrochemical polymerization of N-phenylglycine for amperometric glucose sensing

期刊

EUROPEAN POLYMER JOURNAL
卷 51, 期 -, 页码 130-135

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.eurpolymj.2013.12.005

关键词

Electrochemical polymerization; N-Phenylglycine; Glucose oxidase; Covalent immobilization; Enzyme electrode

资金

  1. Ministry of Education, Culture, Sport, Science and Technology of Japan [23360302]
  2. Grants-in-Aid for Scientific Research [23360302] Funding Source: KAKEN

向作者/读者索取更多资源

N-Phenylglycine, an aniline-derived amino acid, was polymerized electrochemically. The resulting poly(N-phenylglycine) (PPG) was in the form of a thin film having the conductivity of 2.5 x 10(-3) S cm(-1). The surface of the PPG film had a fibrous structure, which made the actual area of the surface larger than the apparent one. Glucose oxidase (GOx) was immobilized covalently on the PPG film by the condensation reaction of amino groups of GOx with the carboxyl groups present on the film. The quantity and activity of the immobilized GOx were determined to be 66.4 mu g cm(-2) and 121 mU cm(-2), respectively. The GOx-immobilized PPG film was applied to amperometric glucose sensing in the presence of p-benzoquinone employed as an electron mediator. It was found that the current response increased with an increase in glucose concentration up to 7.5 mM. The sensitivity was 21.0 mu A mM(-1) cm(-2) with the relative standard deviation of 6.2% (n = 5). (C) 2013 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据