Validation of a Fully Automatic Photoplethysmographic Device for Toe Blood Pressure Measurement

期刊

出版社

W B SAUNDERS CO LTD
DOI: 10.1016/j.ejvs.2010.06.008

关键词

Toe blood pressure; Photoplethysmography; Automation; Validation

资金

  1. ANR (French National Research Agency) [ANR05RNTS01701]

向作者/读者索取更多资源

Objectives: This study was designed to assess the accuracy and reliability of a new, portable, fully automated photoplethysmography (PPG) device for toe blood pressure (TBP) measurement. Design: Within-subject comparison with conventional laser Doppler (LD) measurement. Materials and methods: Four TBP measurements were performed on both lower limbs, alternatively with LD and PPG in 200 patients recruited at the Nimes University Hospital. Reproducibility was assessed by the intraclass correlation coefficient (ICC). The concordance between the two methods was evaluated by Lin's concordance correlation coefficient (CCC), in the whole population as well as in comorbidity subgroups. A potential bias was investigated with the Bland and Altman method. Results: The ICC was 0.887 (95% confidence interval (CI) 0.852-0.913) and 0.893 (0.860-0.918) on the right side (n = 193), 0.905 (0.875-0.928) and 0.898 (0.866-0.922) on the left side (n = 188) for PPG and LD measurements, respectively. The CCC was 0.913 (0.885-0.934) on the right side and 0.915 (0.888-0.937) on the left side, and remained >0.8 regardless of co-morbidities. Conclusions: This new, fully automatic, photoplethysmographic device yielded reliable TBP measurements and showed good agreement with the reference LD system over a wide range of values. (C) 2010 European Society for Vascular Surgery. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据