期刊
EUROPEAN JOURNAL OF RADIOLOGY
卷 68, 期 3, 页码 S49-S53出版社
ELSEVIER IRELAND LTD
DOI: 10.1016/j.ejrad.2008.04.028
关键词
Phase contrast; Animal imaging; Dynamic systems; Phase retrieval
Analyser-based phase contrast X-ray imaging can provide high-contrast images of biological tissues with exquisite sensitivity to the boundaries between tissues. The phase and absorption information can be extracted by processing multiple images acquired at different analyser orientations. Recording both the transmitted and diffracted beams from a thin Laue analyser crystal can make phase retrieval possible for dynamic systems by allowing full field imaging. This technique was used to image the thorax of a mechanically ventilated newborn rabbit pup using a 25 keV beam from the SPring-8 synchrotron radiation facility. The diffracted image was produced from the (1 1 1) planes of a 50 mm x 40 mm, 100 mu m thick Si analyser crystal in the Lane geometry. The beam and analyser were large enough to image the entire chest, making it possible to observe changes in anatomy with high contrast and spatial resolution. (C) 2008 Elsevier Ireland Ltd. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据