Dynamic contour tonometry and Goldmann applanation tonometry: correlation with intracameral assessment of intraocular pressure

期刊

EUROPEAN JOURNAL OF OPHTHALMOLOGY
卷 22, 期 1, 页码 55-62

出版社

WICHTIG EDITORE
DOI: 10.5301/ejo.5000067

关键词

Applanation tonometry; Central corneal thickness; Dynamic contour tonometry; Glaucoma; Intraocular pressure; Manometry

向作者/读者索取更多资源

PURPOSE. To compare intraocular pressure (IOP) measured using a dynamic contour tonometer (DCT) and a Goldmann applanation tonometer (GAT) with in vivo intracameral IOP, and establish the relationship between DCT GAT and central corneal thickness (CCT) in patients with primary open-angle glaucoma (POAG). MATERIALS AND METHODS. We examined 50 eyes of 50 patients with POAG scheduled for glaucoma or cataract surgery. Immediately before surgery, CCT, GAT and DCT IOP were assessed, after which manometry of the anterior chamber was performed. A Bland-Altman plot was used to test the agreement among the 3 measurements of IOP, and univariate and multivariate regression analyses were used to evaluate the effect of CCT on DCT and GAT RESULTS. On average, the DCT readings were 4.0 +/- 1.6 mmHg higher than the GAT readings and 2.3 +/- 2.4 mmHg higher than the manometric readings; the GAT measurements were generally a mean 1.7 +/- 1.8 mmHg lower than the manometric readings. The CCT had an almost similar influence on DCT and GAT measurements (p=0.84). CONCLUSIONS. The DCT-measured IOP was significantly higher than that measured by means of GAT and anterior chamber manometry. The DCT and GAT readings were both influenced by CCT to the same extent.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据