- GACR [13-28882S]
The major features of the selected ion flow tube mass spectrometry (SIFT-MS) analytical method that was conceived and designed for the analysis, in real time, of air obviating sample collections into bags or extraction by pre-concentration of trace compounds onto surfaces are reviewed. The unique analytical capabilities of SIFT-MS for ambient analysis are stressed that allow quantification of volatile organic and inorganic compounds directly from the measurement of physical parameters without the need for regular instrumental calibration using internal or external standards. Then, emphasis is placed on the challenging real-time accurate analysis of single exhalations of humid breath, which is now achieved and readily facilitates wider applications of SIFT-MS in other fields where trace gas analysis has value. The quality of the data obtained by SIFT-MS is illustrated by the quantification of some exhaled breath metabolites that are of immediate relevance to physiology and medicine, including that of hydrogen cyanide in the breath of patients with cystic fibrosis. The current status of SIFT-MS is revealed by a form of a strengths, weakness, opportunities and threats (SWOT) analysis intended to present an objective view of this analytical technique and the likely way forward towards its further development and application.