4.5 Article

Comparison of ELISA and indirect immunofluorescent antibody assay detecting IgM phase II for the diagnosis of acute Q fever

出版社

SPRINGER
DOI: 10.1007/s10096-011-1438-0

关键词

-

向作者/读者索取更多资源

A commercially available enzyme-linked immunosorbent assay (ELISA) detecting phase II-specific IgM for the diagnosis of acute Q fever was compared with indirect immunofluorescent antibody assay (IFA). IFA is the current reference method for the detection of antibodies against , but has disadvantages because the judgment of fluorescence is subjective and tiring, and the test is expensive and automation is not possible. To examine whether phase II IgM ELISA could be used as a screening assay for acute Q fever, we compared the sensitivity and specificity of IFA and ELISA. The sensitivity of the IFA and ELISA tests were 100 and 85.7%, respectively, with a specificity of 95.3 and 97.6%, respectively. Because of the high sensitivity and specificity of the ELISA in combination with the practical disadvantages of the IFA, we introduced a new algorithm to screen samples of patients with symptoms of acute Q fever infection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据