4.6 Article

Adverse results of a decellularized tissue-engineered pulmonary valve in humans assessed with magnetic resonance imaging

期刊

EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY
卷 44, 期 4, 页码 E272-E279

出版社

OXFORD UNIV PRESS INC
DOI: 10.1093/ejcts/ezt328

关键词

Tissue engineering; Pulmonary valve replacement; Immune response; Late gadolinium enhancement; Magnetic resonance imaging

向作者/读者索取更多资源

OBJECTIVES: Matrix P (R) and Matrix P plus (R) tissue-engineered pulmonary valves (TEPV) were offered as an improvement for pulmonary valve replacement (PVR) because of recellularization by host cells. The high frequency of graft failure gave reason to evaluate the underlying morphological substrate using magnetic resonance imaging (MRI) and histology. METHODS: Between June 2006 and August 2008, 17 Matrix P (R) and 10 Matrix P plus (R) TEPVs were implanted in 26 patients with a median age of 12.4 (range: 0.8-38.7, interquartile range: 6.1-18.1) years. The grafts were studied by MRI, and underwent histological examination when explantation was required. RESULTS: Surgical (n = 13) or transcatheter (n = 1) TEPV replacement because of graft failure was needed in 14 cases (52%) 19 (0.5-53) months after implantation. MRI detected significant TEPV stenosis with mild insufficiency (V-max = 3.7 +/- (standard deviation) 0.5 m/s, regurgitant fraction (RGF) = 10 +/- 3%) and stenosis with moderate-to-severe insufficiency (V-max = 3.5 +/- 0.8 m/s, RGF = 38 +/- 10%) in 6 patients, respectively, and severe insufficiency (RGF = 40%) in 1 patient. In patients with graft failure, MRI showed hyperenhancement and TEPV wall thickening. Histology revealed severe inflammation, increased fibrous tissue and foreign-body reaction against valve leaflets and fascial tissue, while TEPV endothelialization was not detected in any case. CONCLUSIONS: The high frequency of Matrix P (R) and Matrix P plus (R) graft failure can be related to inflammation and fibrosis revealed by MRI and histology. Our results do not support the use of these valves for PVR and suggest careful follow-up examinations, including MRI for early detection of graft inflammation and fibrosis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据