期刊
EUROPEAN HEART JOURNAL
卷 31, 期 12, 页码 1470-1476出版社
OXFORD UNIV PRESS
DOI: 10.1093/eurheartj/ehq066
关键词
Incomplete stent apposition; Stent thrombosis; Drug-eluting stent; Restenosis; IVUS; OCT
资金
- Fujita Health University, Toyoake, Japan
- Suzuken Memorial Foundation, Nagoya, Japan
To assess the fate of incomplete stent apposition (ISA) after deployment of sirolimus-eluting stents (SESs). Thirty-two patients having intravascular ultrasound (IVUS)-guided PCI with SESs underwent assessment of stent deployment with quantitative coronary angiography, IVUS, and optical coherence tomography (OCT) pre-procedure, post-procedure, and at 10 months follow-up. Incomplete stent apposition was defined as separation of a stent strut from the inner vessel wall by > 160 mu m. At follow-up, 4.67% of struts with ISA at deployment failed to heal and 7.59% which were well apposed did not develop neointimal hyperplasia even after 10 months. Lesion remodelling was responsible for the development of late ISA in only 0.37% of struts. Failure of adequate neointimal hyperplasia was quantitatively the most important mechanism responsible for persistent acute ISA, classified in previous studies, which relied only on follow-up OCT, as late ISA. Thrombus was visualized in 20.6% of struts with ISA at follow-up and in 2.0% of struts with a good apposition (P < 0.001). In patients with SESs, ISA can fail to heal and even complete apposition can be associated with no neointimal hyperplasia. Incomplete stent apposition without neointimal hyperplasia was significantly associated with the presence of OCT-detected thrombus at follow-up, and may constitute a potent substrate for late stent thrombosis.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据