4.6 Article

Efficacy and safety of implantable cardiac defibrillators for treatment of ventricular arrhythmias in patients with cardiac sarcoidosis

期刊

EUROPACE
卷 15, 期 3, 页码 347-354

出版社

OXFORD UNIV PRESS
DOI: 10.1093/europace/eus316

关键词

Implantable cardiac defibrillator; Cardiac sarcoidosis

资金

  1. Medtronic Inc.
  2. St. Jude Medical
  3. Boston Scientific Corp
  4. Cardiovascular Center at the University of Michigan for development of cardiac sarcoidosis registry
  5. Boston Scientific
  6. Boston Scientific Corp.
  7. Biotronik
  8. CardioInsight Inc.
  9. Thoratec Inc.
  10. Biosense Webster
  11. Biosense Web
  12. Sanofi

向作者/读者索取更多资源

Aims Implantable cardiac defibrillator (ICD) implantation is a class IIA recommendation for patients with cardiac sarcoidosis (CS). However, little is known about the efficacy and safety of ICDs in this population. The goal of this multi-centre retrospective data review was to evaluate the efficacy and safety of ICDs in patients with CS. Methods and results Electrophysiologists at academic medical centres were asked to identify consecutive patients with CS and an ICD. Clinical information, ICD therapy history, and device complications were collected for each patient. Data were collected on 235 patients from 13 institutions, 64.7% male with mean age 55.6 +/- 11.1. Over a mean follow-up of 4.2 +/- 4.0 years, 85 of 234 (36.2%) patients received an appropriate ICD therapy (shocks and/or anti-tachycardia pacing) and 67 of 226 (29.7%) received an appropriate shock. Fifty-seven of 235 patients (24.3%) received a total of 222 inappropriate shocks. Forty-six adverse events occurred in 41 of 235 patients (17.4%). Patients who received appropriate ICD therapies were more likely to be male (73.8 vs. 59.6%, P = 0.0330), have a history of syncope (40.5 vs. 22.5%, P = 0.0044), lower left ventricular ejection fraction (38.1 +/- 15.2 vs. 48.8 +/- 14.7%, P <= 0.0001), ventricular pacing on baseline electrocardiogram (16.1 vs. 2.1%, P = 0.0002), and a secondary prevention indication (60.7 vs. 24.5%, P < 0.0001) compared with those who did not receive appropriate ICD therapies. Conclusion Patients with CS and ICDs are at high risk for ventricular arrhythmias. This population also has high rates of inappropriate shocks and device complications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据