Assessment of gingival thickness using an ultrasonic dental system prototype: A comparison to traditional methods

期刊

ANNALS OF ANATOMY-ANATOMISCHER ANZEIGER
卷 199, 期 -, 页码 98-103

出版社

ELSEVIER GMBH, URBAN & FISCHER VERLAG
DOI: 10.1016/j.aanat.2014.02.010

关键词

Gingival thickness; Biotype assessment; High frequency ultrasound; Trans-gingival probing

向作者/读者索取更多资源

Knowledge of periodontal anatomy is essential when performing surgical and non-surgical procedures in the field of oral healthcare. Gingival thickness (GT) is often assessed for this purpose. A dental system prototype was recently developed for quantitative, non-invasive GT assessment by high-frequency (HF) ultrasound. Laboratory trials were conducted to validate system performance against a traditional method of assessment. A system with a 50 MHz broadband, spherically-focused transducer was used. The transducer was housed in a small, hand-held probe equipped with a continuous water supply. A-scans were obtained and thickness at each location was determined. For comparison, the traditional method of transgingival probing through tissue with an endodontic k-file needle was also implemented. Preliminary experiments were performed on phantoms simulating the anatomical and acoustic properties of human periodontal tissues. A porcine cadaver was obtained for further laboratory trials. The speed of sound through porcine gingiva was determined to be 1564 +/- 21 m/s. Finally, a multiple-point experiment involved GT assessment in an array of locations on the buccal gingival surface in the fourth quadrant. Ultrasonic measurements were found to yield similar GT values to those obtained from invasive methods. Results obtained in this experiment validate the applicability of ultrasound as a diagnostic tool for assessing periodontal anatomy. (C) 2014 Elsevier GmbH. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据