4.5 Article

Improving MCE with electrochemical detection using a bubble cell and sample stacking techniques

期刊

ELECTROPHORESIS
卷 30, 期 19, 页码 3339-3346

出版社

WILEY
DOI: 10.1002/elps.200900316

关键词

Bubble cell; Electrochemical detection; Field amplified sample injection; Field amplified sample stacking; MCE

资金

  1. National Institute of Health [9 EB 004876-01AI]

向作者/读者索取更多资源

Two efforts to improve the sensitivity and limits of detection for MCE with electrochemical detection are presented here. One is the implementation of a capillary expansion (bubble cell) at the detection zone to increase the exposed working electrode surface area. Bubble cell widths were varied from 1 x to 10 x the separation channel width (50 mu m) to investigate the effects of electrode surface area on detection sensitivity, LOD, and separation efficiency. Improved detection sensitivity and decreased detection limits were obtained with increased bubble cell width, and LODs of dopamine and catechol detected in a 5 x bubble cell were 25 and 50 nM, respectively. Meanwhile, fluorescent imaging results demonstrated similar to 8 and similar to 12% loss in separation efficiency in 4 x and 5 x bubble cell, respectively. Another effort at reducing the LOD involves using field amplified sample injection for gated injection and field amplified sample stacking for hydrodynamic injection. Stacking effects are shown for both methods using amperometric detection and pulsed amperometric detection. The LODs of dopamine in a 4 x bubble cell were 8 and 20 nM using field amplified sample injection and field amplified sample stacking, respectively. However, improved LODs were not obtained for anionic analytes using either stacking technique.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据