4.5 Article

Automation of nanoflow liquid chromatography-tandem mass spectrometry for proteome and peptide profiling analysis by using a monolithic analytical capillary column

期刊

ELECTROPHORESIS
卷 29, 期 8, 页码 1612-1618

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elps.200700513

关键词

automation; monolithic analytical capillary column; nano-LC-MS/MS; peptide separation; sample introduction

向作者/读者索取更多资源

An automated nano-LC-MS/MS platform without trap column was established, which only used a 20 cm lauryl methacrylate-ethylene dimethacrylate (LMA-EDMA) monolithic capillary column to allow preconcentration and separation of peptides. The monolithic column had the advantages of good permeability and low backpressure resulting in higher flow rates for capillary columns. Tryptic digests of bovine albumin and yeast protein extract were tested using the monolithic column system. High proteomic coverage using this approach were demonstrated in this study. Furthermore, peptide samples extracted from mouse liver were separated by using the monolithic column system combined with size-exclusion chromatography prefractionation. This monolithic column system might be a promising alternative for the automated system previously using a trap column for routine proteome and peptide profiling analysis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据