4.6 Article

Ion selective electrodes for protonable nitrogen containing analytes: Metallacarboranes as active membrane components

期刊

ELECTROCHIMICA ACTA
卷 113, 期 -, 页码 94-98

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.electacta.2013.09.025

关键词

Ion selective electrodes; Borane anion; Nitrogenous bioactive compounds; Metallacarborane

资金

  1. Austrian Research Promotion Agency (Osterreichische Forschungsforderungsgesellschaft, FFG)
  2. Government of Lower Austria
  3. Generalitat de Catalunya [2009/SGR/00279]
  4. Spanish Ministerio de Economia y Competitividad [CTQ2010-16237]

向作者/读者索取更多资源

Ion selective electrodes with simple electroactive membrane components have been prepared for the testing of dopamine, nicotinic acid, nicotinamide, histamine and metformin in aqueous solutions. The electroactive membrane component is the salt made of the protonated targeted analyte and the metallacarborane anion [Co(1,2-C2B9H11)(2)](-). The reason for the good performance of [Co(1,2-C2B9H11)(2)](-) is its large capacity to produce weak hydrogen and dihydrogen bonds with the cation and the plasticizer that prevents leakage of the electroactive component. In this paper, we prove how easy is to produce a sensing material for a specific target. The only requirement is that the target has a protonable nitrogen, a situation encountered in many biological molecules or in many bio-relevant molecules. The analytical parameters for evaluation of ISEs performances, such as linearity range, slope, limit of detection, response time, and selectivity were studied and determined and all the produced selective ISEs have given favorable answer. The proposed electrodes showed Nernstian responses for protonable nitrogen containing analytes for almost 4 decades of concentrations, 1.00 x 10(-6) to 1.00 x 10(-2) mol dm(-3), with fast response time (5s) and adequate operational lifetime (50 days). For Dopamine the prepared ISEs exhibited linear responses with slope of ca. 58 mV decade(-1) within the range 5.00 x 10(-6) to 1.00 x 10(-2) mol dm(-3) and with a low detection limit of 1 mu mol dm(-3). The limits of detection ranged from 0.7 mu mol dm(-3) for nicotinamide up to 6 mu mol dm(-3) for metformin. (C) 2013 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据