期刊
ELECTROANALYSIS
卷 26, 期 4, 页码 864-872出版社
WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.201300590
关键词
Nitrate-selective electrode; Solid contact; Ionic liquid; Potentiometry
A simple, fast and cheap method of preparation of solid contact nitrate ion-selective electrode is proposed. The electrode membrane phase consist of only three components: PVC, plasticizer and ionic liquid (IL).The ionic liquid trihexyltetradecylphosphonium chloride is used in triple function as ionophore, as lipophilic ionic component in order to reduce membrane resistance, and as transducer media in order to stabilize the potential of internal Ag/AgCl electrode. The electrical properties of the membrane were studied by electrochemical impedance spectroscopy and the influence of the interfacial water film was evaluated by potentiometric water layer test.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据