4.5 Article

Trace Levels Determination of Ammonium by Flow Injection Analysis Using Gas-Diffusion and Capacitively Coupled Contactless Conductivity Detection

期刊

ELECTROANALYSIS
卷 23, 期 11, 页码 2594-2600

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.201100317

关键词

Gas-diffusion; Contactless conductivity detection; Ammonium determination; Rainwater

资金

  1. CNPq (Conselho Nacional de Desenvolvimento Cientifico e Tecnologico)
  2. FAPESP (Fundacao de Amparo a Pesquisa do Estado de Sao Paulo) [2008/53868-0, 2008/57805-2]
  3. Mackpesquisa

向作者/读者索取更多资源

A sensitive analytical method based on gas-diffusion flow injection system for ammonium ion determination at trace levels with capacitively coupled contactless conductivity detection ((CD)-D-4) is described. The effects of the chemical composition of the acceptor stream, operating frequency of the conductivity detector, acceptor and donor flow rates, injected volume, and some potential interferents were evaluated. By injecting 100 mu L samples, the analytical method showed linear response for ammonium from 2.0 up to 100 mu mol L-1 and a limit of detection of 0.8 mu mol L-1 (14 ppb). At a flow rate of 0.10 mL min(-1) for the acceptor solution, an analytical throughput of 35 injections per hour was achieved. The new flow injection system was successfully applied for ammonium determination in rainwater from Sao Paulo city.