4.5 Article

Flow Injection Analysis of Ethambutol in Antituberculosis Drugs Using a Graphite-Paraffin Electrode as Amperometric Detector

期刊

ELECTROANALYSIS
卷 23, 期 11, 页码 2582-2585

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.201100195

关键词

Antituberculosis drugs; Carbon-based composite electrode; Ethambutol; FIA

资金

  1. FAPEMIG (Fundacao de Amparo a Pesquisa do Estado de Minas Gerais)
  2. CNPq (Conselho Nacional de Desenvolvimento Cientifico e Tecnologico)
  3. CAPES (Coordenacao de Aperfeicoamento de Pessoal de Nivel Superior)
  4. Propesq-UFJF (Pro-reitoria de Pesquisa da Universidade Federal de Juiz de Fora)

向作者/读者索取更多资源

A graphite-paraffin composite electrode has been used for the determination of ethambutol, an antituberculosis drug, in pharmaceutical samples by anodic oxidation. Voltammetric studies were used to characterise the electro-chemical behaviour of the ethambutol in different pH solutions (3.0, 8.0 and 12.0). Quantification was carried out using amperometric detection. The detector was assembled in a flow injection apparatus and operated at + 1.1 V (vs. Ag/ AgCl(NaCl, sat)). The conditions for the best electrode's response were determined by studying the influence of sample volume and flow rate. The linear response for the method was extended up to a 1.5 mmolL(-1) ethambutol solution with a detection limit of 0.1 mmolL(-1). The repeatability for successive injections of 1.0 mmolL(-1) ethambutol solution was 0.9% (n = 26), demonstrating a good performance of the developed amperometric sensor. The analytical frequency was calculated as approximately 47 determinations h(-1). Two different commercial samples were successfully analysed and the results were in good agreement with those obtained by using spectrophotometry. The advantages of the proposed amperometric method over the reference method include simplicity, low consumption of reagents and increased analytical frequency.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据