4.5 Article

Usage of Bismuth Film Electrode as Biosensor Transducer for Alkaline Phosphatase Assay

期刊

ELECTROANALYSIS
卷 22, 期 13, 页码 1519-1523

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/elan.200900447

关键词

Bismuth film electrode; Alkaline phosphatase (ALP); Biosensors; Bismuth

向作者/读者索取更多资源

A biosensor for monitoring alkaline phosphatase (ALP) enzyme activity was developed by using bismuth film electrode (BiFE) as transducer. The aim was to test the possible applications of BiFE as transducer in this biosensing platform. ALP was immobilized onto the BiFE electrode by means of gelatin which was then crosslinked by glutaraldehyde. The measurement was based on monitoring of p-nitrophenol which was released after enzymatic reaction of p-nitrophenyl phosphate (pNPP) substrate. The effect of enzyme activity (in Unit) was examined for 100 mu M of pNPP substrate and a good linearity was observed between 1.28 and 4.26 U activities with the equation of y = 0.214x - 0.201 and R-2 = 0.998 where x and y show enzyme activity in unit and current response in mu A respectively. Analytical characteristics were also examined after the optimization of experimental parameters.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据