期刊
DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE
卷 69, 期 3, 页码 240-244出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/j.diagmicrobio.2010.09.023
关键词
Commercial kits; DNA extraction; DNA purification; Gastrointestinal pathogens; Stool
In this study, 5 different commercial DNA extraction systems were tested on a stool sample collection containing 81 clinical stool specimens that were culture-positive for diarrheagenic Escherichia coli, Campylobacter jejuni, Salmonella enterica, or Clostridium difficile. The purified DNAs were analyzed by polymerase chain reaction (PCR) directed toward the relevant organisms. The results showed that conventional PCR combined with the extraction systems BioRobot EZI (Qiagen, Hilden, Germany), Bugs'n Beads (Genpoint, Oslo, Norway), ChargeSwitch (Invitrogen, Paisley, UK), QIAamp Stool Mini Kit (Qiagen), and 2 protocols (generic and Specific A) for EasyMag (BioMerieux, Marcy I'Etoilc, France) were able to identify 89%, 62%, 85%, 88%, 85%, and 91%, respectively, of the pathogens originally identified by conventional culture-based methods. When TaqMan PCR was combined with the EasyMag Specific A protocol, 99% of the samples were correctly identified. The results demonstrate that the extraction efficiencies can vary significantly among different extraction systems, careful optimization may have a significant positive effect, and the use of sensitive and specific detection methods like TaqMan PCR is an ideal choice for this type of analysis. (C) 2011 Elsevier Inc. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐