A glucose meter accuracy and precision comparison: The FreeStyle flash versus the Accu-Chek advantage, Accu-Chek Compact Plus, Ascensia Contour, and the BD Logic

期刊

DIABETES TECHNOLOGY & THERAPEUTICS
卷 10, 期 2, 页码 102-110

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2007.0244

关键词

-

向作者/读者索取更多资源

Background: This study compared the accuracy and precision of five blood glucose (BG) meters. Methods: Diabetes patients undergoing venipuncture for glucose testing were randomized to one of two groups consisting of three meters: FreeStyle (R) Flash (Abbott Diabetes Care, Alameda, CA), Accu-Chek (R) Advantage (Roche Diagnostics Corp., Indianapolis, IN), and Accu-Chek Compact Plus (Roche Diagnostics) or FreeStyle Flash, Ascensia (R) Contour (R) (Bayer Healthcare, Diagnostic Division, Tarrytown, NY), and BD Logic (R) (BD Diabetes Care, Franklin Lake, NJ). Within 5 min following venipuncture, duplicate finger BG measurements from three ipsilateral fingers were taken. Finger glucose measurements were compared with laboratory reference values. Accuracy was assessed by a Clarke error grid analysis (EGA) and within 10% of the laboratory value criteria. Meter precision was determined by calculating the absolute mean differences in glucose values between duplicate samples. Results: Finger sticks were obtained from 202 patients. Mean venipuncture BG was 148 mg/dL (SD +/- mg/64 dL; range 25-439 mg/dL). Accuracy by Clarke EGA (Zone A results) was demonstrated in 69% of Advantage samples, 75% of Compact Plus, and 96% of the first group of Flash versus 88% of the Contour, 67% of the Logic, and 91% of the second Flash samples (P < 0.05 for both Flash and Contour). Meter accuracy using the 10% criteria was demonstrated in 30%, 38%, 70%, 46%, 48%, and 68% of the samples, respectively (P < 0.05 for both Flash groups compared to each of the other meters). There were no differences in meter precision. Conclusions: No statistically significant differences in accuracy were evident using the Clarke EGA criteria (pooled results of Zone A and B), though the more strict 20% accuracy criteria (Zone A results only) found the Flash and Contour to have significantly greater accuracy compared to the Advantage, Compact Plus, and the Logic. Using the 10% accuracy criteria found the Flash to have significantly greater accuracy than each of the other four meters. All five meters demonstrated similar precision.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据