4.6 Article

Influence of loading frequency on implant failure under cyclic fatigue conditions

期刊

DENTAL MATERIALS
卷 25, 期 11, 页码 1426-1432

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.dental.2009.06.015

关键词

Dental implant; Fatigue testing; Strain rate; Strain; Loading frequency

资金

  1. ITI Foundation for the Promotion of Oral Implantology, Switzerland

向作者/读者索取更多资源

Objectives. Preliminary studies on implant fatigue testing suggested that fractures were more likely to occur at 2Hz than at 30 Hz (chi(2), p < 0.05). This investigation explores frequency and base elastic modulus effects on strain, strain rate and failure. Methods. A total of 66 implants were mounted in different base materials (acrylic, glass. filled epoxy, aluminum) and loaded up to 106 cycles per ISO 14801 (20N to 420-500 N) at frequencies of 2Hz and 30 Hz (chosen to accelerate frequency as the stressor). Absolute strain magnitudes and strain rates under varying loading conditions and with different base materials were measured using one strain-gauged implant. Failure probability distributions were analyzed by both Weibull and life data analysis. Measured strain was used to validate an FEA model. Fracture surfaces were examined by SEM. Results. Number of failures and failure-rates-per-cycle differed significantly between implants tested at 2Hz versus 30 Hz (p < 0.05). Strain magnitude was independent of frequency. Strain rates were highly correlated with frequency (linear r(2) > 0.99) and differed significantly under failure conditions (420 N): 2Hz = 8292 mu strain/s; (500 N): 30 Hz = 80,840 mu strain/s. Measured and FEA-calculated strains were similar. Fracture surfaces were indistinguishable (2Hz versus 30 Hz). Significance. Fatigue failure was significantly more likely at 2 Hz than 30 Hz whereas base material and loading magnitude seemed to have only minor influence. Absolute strain was identical at these frequencies suggesting strain rate sensitivity for this commercially pure titanium implant. Both the Weibull and SEM analyses support an identical failure mechanism with damage accumulation more severe at lower frequencies, an interpretation consistent with strain rate sensitivity. (C) 2009 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据