4.4 Article

Potentiometric urea biosensor based on poly(glycidylmethacrylate)-grafted iron oxide nanoparticles

期刊

CURRENT APPLIED PHYSICS
卷 13, 期 1, 页码 280-286

出版社

ELSEVIER
DOI: 10.1016/j.cap.2012.07.025

关键词

Urease; Potentiometry; Magnetic nanoparticle; Poly(glycidylmethacrylate); Biosensor

资金

  1. DPT (State Planning Center)
  2. Fatih University Research Support Office [P50021104-B]

向作者/读者索取更多资源

Urease enzyme was covalently attached on the poly(glycidylmethacrylate) (PGMA)-grafted iron oxide nanoparticles on Au electrode for the fabrication of urea biosensor. The telomere of poly(glycidylmethacrylate) (PGMA) with a trimethoxysilyl terminal group was synthesized by telomerization of glycidylmethacrylate. Iron oxide nanoparticles were coated with telomere of poly(glycidylmethacrylate) in order to obtain good enzyme immobilization platform. The telomere and nanoparticles were characterized by using Fourier transform infrared (FT-IR), X-ray diffraction (XRD), transmission electron microscopy (TEM) and thermal gravimetric analysis (TGA). The biosensor's potentiometric response was measured as a function of urea concentration in phosphate buffer solution (10 mM, pH 7.5) and showed a linear range of 0.25-5.0 mM urea. The produced biosensor exhibited a good response time of similar to 8 s and was stable for about two months. The basic features (optimum pH, optimum temperature, interference and storage stability) of the enzyme electrode were determined. (C) 2012 Elsevier B.V. All rights reserved.