期刊
CURRENT ANALYTICAL CHEMISTRY
卷 6, 期 3, 页码 203-208出版社
BENTHAM SCIENCE PUBL LTD
DOI: 10.2174/157341110791517025
关键词
Modified glassy carbon; Indole-3-carboxaldehyde film; Dopamine; Ascorbic acid; Uric acid; Simultaneous determination
资金
- TUBITAK (Scientific and Technological Research Council of Turkey) [106T622]
Indole-3-carboxaldehyde (In3C) molecules were deposited on the glassy carbon (GC) electrode by electrochemical methods in acetonitrile (containing 0.05 M TBATFB). Deposition of In3C was performed by cyclic voltammetry scanning from 0.0 V to 2.0 V at a scan rate of 50 mV s(-1) for 10 cycles in 1 mM In3C solution. The modified surface (In3C-GC) was characterized by cyclic voltammetric response of ferrocene as a redox probe. Elemental composition of the surface was analyzed by X-ray photoelectron spectroscopy (XPS). The applicability of this modified surface was then investigated for the determination of dopamine (DA) and uric acid (UA) in the presence of ascorbic acid (AA) using differential pulse voltammetry. Detection limits were calculated as 1.70 mu M for DA and 4.99 mu M for UA. The linear ranges of the calibration curves were found between 10 mu M -100 mu M for both DA and UA in the presence of 100 mu M AA. The proposed method was applied to the determination of DA and UA in pharmaceutical preparations and urine samples with satisfactory results.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据