4.4 Article

Quantitative 3-Dimensional Corneal Imaging In Vivo Using a Modified HRT-RCM Confocal Microscope

期刊

CORNEA
卷 32, 期 4, 页码 E36-E43

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/ICO.0b013e31825ec44e

关键词

confocal microscopy; cornea; imaging; 3-D reconstruction

资金

  1. NIH [R01 EY013322, P30 EY020799]
  2. Research to Prevent Blindness, Inc., New York, NY

向作者/读者索取更多资源

Purpose: The purpose of this study was to develop and test hardware and software modifications to allow quantitative full-thickness corneal imaging using the Heidelberg Retina Tomograph (HRT) Rostock Corneal Module. Methods: A personal computer-controlled motor drive with positional feedback was integrated into the system to allow automated focusing through the entire cornea. The left eyes of 10 New Zealand white rabbits were scanned from endothelium to epithelium. Image sequences were read into a custom-developed program for depth calculation and measurement of sublayer thicknesses. Three-dimensional visualizations were also generated using Imaris. In 6 rabbits, stack images were registered, and depth-dependent counts of keratocyte nuclei were made using Metamorph. Results: The mean epithelial and corneal thickness measured in the rabbit were 47 +/- 5 mu m and 373 +/- 25 mu m, respectively (n = 10 corneas); coefficients of variation for repeated scans were 8.2% and 2.1%. Corneal thickness measured using ultrasonic pachymetry was 374 + 17 mu m. The mean overall keratocyte density measured in the rabbit was 43,246 +/- 5603 cells per cubic millimeter in vivo (n = 6 corneas). There was a gradual decrease in keratocyte density from the anterior to posterior cornea (R = 0.99), consistent with previous data generated in vitro. Conclusion: This modified system allows high-resolution 3-dimensional image stacks to be collected from the full-thickness rabbit cornea in vivo. These data sets can be used for interactive visualization of corneal cell layers, measurement of sublayer thickness, and depth-dependent keratocyte density measurements. Overall, the modifications significantly expand the potential quantitative research applications of the HRT Rostock Cornea Module microscope.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据