期刊
COLORECTAL DISEASE
卷 16, 期 12, 页码 O435-O439出版社
WILEY-BLACKWELL
DOI: 10.1111/codi.12742
关键词
Compression; anastomosis; leakage
AimCompression anastomosis has proved to be safe for rectal reconstruction with leak rates comparable to those observed using circular stapling devices. However, there are no data on whether the metallic compression ring alters the ease of diagnosis or the treatment in cases of leakage. In this study, we present our experience with early leakage following compression anastomosis. MethodA prospective registry was used for data review. Patients with anastomotic leakage following compression anastomosis between November 2008 and September 2013 were included. ResultsIn all, 197 (92 female) patients were operated using a novel compression device. Early leakage was found in 10 (5.1%) patients after a median of 5 (3-14) days. The radiologist was able to detect leakage using CT in nine out of 10 cases unequivocally. Removal of the ring was necessary in eight of the 10 cases, and salvage of the anastomosis was feasible on six occasions. In all diverted cases with a low anastomosis, a transanal repair of the defect was feasible in three cases, including a single patient with complete separation of the anastomosis. ConclusionArtefacts on the CT scan caused by the compression ring did not hamper the diagnosis of anastomotic leakage. Removal of the ring in the early postoperative period is not associated with complete separation of the bowel ends. Salvage of anastomosis is feasible in most cases.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据