4.6 Article Proceedings Paper

High Cup Angle and Microseparation Increase the Wear of Hip Surface Replacements

期刊

CLINICAL ORTHOPAEDICS AND RELATED RESEARCH
卷 467, 期 9, 页码 2259-2265

出版社

SPRINGER
DOI: 10.1007/s11999-009-0830-x

关键词

-

资金

  1. MRC [G0701221] Funding Source: UKRI
  2. Arthritis Research UK Funding Source: Medline
  3. Medical Research Council [G0701221] Funding Source: Medline
  4. Medical Research Council [G0701221] Funding Source: researchfish

向作者/读者索取更多资源

High wear rates and high patient ion levels have been associated with high (> 55A degrees) cup inclination angles for metal-on-metal surface replacements. Wear rates and patterns have been simulated for ceramic-on-ceramic bearings by applying microseparation to replicate head offset deficiency. We tested 39-mm metal-on-metal surface replacements (n = 5) in a hip simulator with (A) an increased cup inclination angle of 60A degrees and (B) an increased cup inclination angle and microseparation over 2 million cycles. (A) resulted in a ninefold increase in wear rate and (B) resulted in a 17-fold increase in wear rate compared to a standard gait condition study. Wear particles produced under microseparation conditions were larger than those produced under standard conditions but of similar shape (round to oval). The data suggest both head and cup position influence the wear of surface replacements; we believe it likely bearings with high wear either have a high cup inclination angle, an offset deficient head, or a combination of both.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据