4.5 Article

A study of the bone healing kinetics of plateau versus screw root design titanium dental implants

期刊

CLINICAL ORAL IMPLANTS RESEARCH
卷 20, 期 3, 页码 232-239

出版社

WILEY
DOI: 10.1111/j.1600-0501.2008.01640.x

关键词

endosseous implants; implant design; osseointegration

资金

  1. Bicon<SUP>(R)</SUP> (Boston, USA
  2. Implant Company

向作者/读者索取更多资源

This study was designed to compare the bone healing process around plateau root from (PRF) and screw root from (SRF) titanium dental implants over the immediate 12 week healing period post implant placement. 32PRF and 32SRF implants were placed in 8 beagle dogs at 12, 8, 5 and 3 weeks prior to enthanisation using a bilaterally balanced distribution. Undecalcified ground sections were prepared from the biopsies taken and histometric measurements of bone implant contact (BIC) and bone area fraction occupancy (BAFO) were made on the middle 5 mm portion of each 8 mm implant root length. The analysis showed that although measurements of bone to implant contact (BIC) and bone area fraction occupancy (BAFO) tended to be greater for the SRF implants at all four time points, the differences in measurements between implant types did not reach statistical significance (P=0.07, P=0.06). The effect of time on BIC and BAFO was found to be strongly significant for both implant types thus indicating a statistically significant increase in BIC and BAFO overall with time (P=0.004, P=0.002). Furthermore, both PRF and SRF implants behaved similarly over time with measurements of BIC and BAFO progressing in parallel. Histomorphologic analysis of these sections demonstrated the prominent role of woven bone (callus) in the bone healing process around PRF implants. The results can be interpreted to indicate a comparable development of secondary stability for both PRF and SRF implant designs. However, as these parameters reflect the structural connection between implant and bone and not the functional properties of the bone to implant interface, they cannot be regarded as comprehensive measures of osseointegration. This particularly relevant given the reduced load bearing capacity of woven bone. To cite this article:Leonard G, Coelho P, Polyzois I, Stassen L, Claffey N. A study of the bone healing kinetics of plateau versus screw root design titanium dental implants.Clin. Oral Impl. Res. 20, 2009; 232-239.doi: 10.1111/j.1600-0501.2008.01640.x.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据