4.7 Article

Antimicrobial susceptibility testing in 90min by bacterial cell count monitoring

期刊

CLINICAL MICROBIOLOGY AND INFECTION
卷 19, 期 3, 页码 286-291

出版社

WILEY-BLACKWELL
DOI: 10.1111/j.1469-0691.2012.03800.x

关键词

Antibiotics; cell count monitoring; flow-cytometry; microbiology; susceptibility testing

向作者/读者索取更多资源

Clin Microbiol Infect 2013; 19: 286291 Abstract The rise in antimicrobial resistance has become a serious global health problem. Restrictive use of antibiotics seems the only option to temper this accession since research in new antibiotics has halted. Antimicrobial stewardship programmes rely on quick access to susceptibility data. This study evaluated the concept of bacterial cell count monitoring as a fast method to determine susceptibility. Escherichia coli, Pseudomonas aeruginosa and Staphylococcus aureus strains were tested for amoxicillin/piperacillin and gentamicin by three conventional methods (VITEK2 (R), Etest (R) and broth-macrodilution). Bacterial cell count monitoring reliably predicted susceptibility after 90min for Escherichia coli and after 120min for Pseudomonas aeruginosa and Staphylococcus aureus without any minor, major or very major discrepancies. Time-to-result was reduced by 74%, 83% and 76%, respectively. Bacterial cell count monitoring shows great potential for rapid susceptibility testing.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据