Evaluation of the Abbott Real-Time HIV-1 quantitative assay with dried blood spot specimens

期刊

CLINICAL MICROBIOLOGY AND INFECTION
卷 15, 期 1, 页码 93-97

出版社

ELSEVIER SCI LTD
DOI: 10.1111/j.1469-0691.2008.02116.x

关键词

Antiretroviral therapy; dried blood spot; human immunodeficiency virus; monitoring; viral load assay

资金

  1. VI Programma Nazionale di Ricerca AIDS 2006, Italian Ministry of Health [30G. 58]
  2. University of Siena, Italy

向作者/读者索取更多资源

The Abbott Real-Time HIV-1 assay was evaluated for its performance in quantification of human immunodeficiency virus type 1 (HIV-1) RNA in dried blood spot (DBS) samples. In total, 169 blood samples with detectable plasma HIV-1 RNA were used to extract RNA from paired DBS and liquid plasma samples, using the automated Abbott m Sample Preparation System (m2000sp). HIV-1 RNA was then quantitated by the m2000rt RealTime analyser. RNA samples suitable for real-time PCR were obtained from all but one (99.4%) of the DBS samples and HIV-1 RNA was detected in 163/168 (97.0%) samples. The correlation between HIV-1 RNA values measured in paired DBS and plasma samples was very high (r = 0.882), with 78.5% and 99.4% of cases differing by < 0.5 and 1.0 log, respectively. Retesting of DBS replicates following 6 months of storage at 2-8 degrees C showed no loss of HIV-1 RNA in a subset of 89 samples. The feasibility of DBS testing coupled with automated sample processing, and the use of a latest-generation FDA-approved real-time PCR-based system, represents an encouraging first step for viral load measurement in reference centres in developing countries where access to antiretroviral therapy is expanding.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据