期刊
CLINICAL INFECTIOUS DISEASES
卷 49, 期 7, 页码 1090-1093出版社
OXFORD UNIV PRESS INC
DOI: 10.1086/644743
关键词
-
We found that the sensitivities of 3 rapid influenza antigen tests for pandemic influenza A/H1N1 virus were low to moderate: BD Directigen EZ Flu A+B test (Becton Dickinson), 46.7%; BinaxNOW Influenza A& B (Inverness Medical), 38.3%; and QuickVue Influenza A+B Test (Quidel), 53.3%. A patient with influenza-like illness who has a negative rapid antigen test result should undergo further testing using reverse-transcription polymerase chain reaction.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐