Primary Stability of a Hybrid Implant Compared with Tapered and Cylindrical Implants in an Ex Vivo Model

期刊

出版社

WILEY
DOI: 10.1111/cid.12205

关键词

animal model; dental implant; implant stability quotient; maximum insertion torque; primary stability; push-out test

向作者/读者索取更多资源

BackgroundHybrid implants are thought to offer good stability and bone-implant contact. PurposeThis study sought to evaluate the primary stability of a hybrid implant compared with tapered and cylindrical implants in an ex vivo model and compare undersized drilling with standard drilling in attaining primary stability. Materials and MethodsHybrid (Straumann (R) Bone Level; BL), tapered (Straumann (R) Tapered Effect; TE), and cylindrical (Straumann (R) Standard Plus; SP) implants were inserted into 15 mm-thick porcine iliac crest blocks using standard (corticocancellous or cancellous bone) or undersized (cancellous bone only) drilling (n=7). Stability was evaluated using maximum insertion torque (MIT), implant stability quotient (ISQ), and push-out tests. ResultsThe mean MIT for BL and TE was significantly higher than that for SP (p<0.05). With standard drilling in corticocancellous bone, the mean ISQ for TE was significantly higher than that for SP (p<.05). The mean MIT, ISQ, and push-out test scores for BL, TE, and SP with standard drilling in corticocancellous bone were significantly higher than those with undersized or standard drilling in cancellous bone (p<.05). ConclusionsIn this ex vivo model, the hybrid implant demonstrated promising primary stability compared with the cylindrical implant. The improved stability of the hybrid implant was especially evident in cancellous bone. The influence of the cortical layer should be recognized when selecting implants for primary stability.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据