How Accurate Is CBCT in Measuring Bone Density? A Comparative CBCT-CT In Vitro Study

期刊

出版社

WILEY
DOI: 10.1111/cid.12027

关键词

bone density; computed tomography; cone beam; Hounsfield value

向作者/读者索取更多资源

Purpose: Recently, cone beam computed tomography (CBCT) has become widely used for oral and maxillofacial imaging. Twenty dry mandibles were CBCT and conventional multislice CT scanned to evaluate if there is a statistically significant difference between the bone density values they produce, defined as gray density values, and to determine any correlation between them. Materials and Methods: Using software and a radiographic template, the CT and CBCT scan images were overlapped, and two data sets were created, each one giving the respective gray values (voxel value [ VV] or Hounsfield unit [ HU]) of the same area with the same spatial coordinates. For the statistical analysis, t-test, Pearson's correlation, and Pearson's r were used. Results: The differences between the CBCT (VV) and CT (HU) gray density values were statistically significant (p <= 2.05), whereas the Pearson's correlation coefficients and Pearson's r-values demonstrated a statistically significant linear correlation between VV and HU gray density values. Conclusion: The lower radiation dose and reduced costs of CBCT make this a useful substitute for CT; however, this study has shown that, in order to more accurately define the bone density with CBCT, a conversion ratio needs to be applied to the VV.