Diagnosing Diabetes Mellitus: Performance of Hemoglobin A1c Point-of-Care Instruments in General Practice Offices

期刊

CLINICAL CHEMISTRY
卷 59, 期 12, 页码 1790-1801

出版社

OXFORD UNIV PRESS INC
DOI: 10.1373/clinchem.2013.210781

关键词

-

向作者/读者索取更多资源

BACKGROUND: Hemoglobin A(1c) (Hb A(1c)) measurement by hospital laboratory instruments, but not by point-of- care (POC) instruments, has been recommended for use to diagnose diabetes mellitus. We evaluated results from 13 Hb A(1c) external quality assurance (EQA) surveys over a 6-year period in Norway, from both POC instruments used in general practice (GP) offices and instruments in hospital laboratories, against the analytical quality specifications recommended for use of Hb A(1c) to diagnose diabetes mellitus. METHODS: All GP offices (n = 1288) and hospital laboratories (n = 52) measuring Hb A(1c) in Norway participated in the EQA survey. The percentage of participants that performed measurements within the quality specifications was calculated. Pooled within-laboratory CVs were estimated for the Afinion, DCA 2000, DCA 2000+, DCA Vantage (TM), and Nycocard Hb A(1c) Reader instruments and for hospital laboratory instruments. RESULTS: Between 60% to 90% of Afinion and DCA users and hospital laboratories performed Hb A(1c) measurements within the quality specifications for both trueness (6.0%) and imprecision (CV <= 2.0%) at 2 levels in each EQA survey. The pooled within-laboratory CVs for the Afinion and DCA instruments and hospital laboratories were below the recommended limit of 2.0% for most of the surveys. CONCLUSIONS: A large proportion of GP offices using Afinion and DCA POC instruments to measure Hb A(1c) fulfill the analytical quality specifications for diagnosing diabetes mellitus, and these instruments demonstrate analytical quality comparable to that of hospital laboratory instruments. When GP offices participate in a stringent quality assurance program and generate Hb A(1c) measurements that meet analytical quality specifications, these measurements can be recommended for use to diagnose diabetes mellitus. (C) 2013 American Association for Clinical Chemistry

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据