4.7 Article

Prime Time for Enzymatic Creatinine Methods in Pediatrics

期刊

CLINICAL CHEMISTRY
卷 55, 期 3, 页码 549-558

出版社

AMER ASSOC CLINICAL CHEMISTRY
DOI: 10.1373/clinchem.2008.116863

关键词

-

资金

  1. Abbott Diagnostics (Irving, Texas)
  2. Roche Diagnostics (Penzberg, Germany)
  3. Siemens Healthcare Diagnostics (Tarrytown, New York)

向作者/读者索取更多资源

BACKGROUND: The literature regarding the nonspecificity of applications of the Jaffe (alkaline picrate) reaction for creatinine is generally outdated. We conducted a specificity study to update the nonspecificity information for current Jaffe and enzymatic creatinine assays. METHODS: Two serum pools with creatinine concentrations within the pediatric reference interval were spiked with albumin, IgG, unconjugated bilirubin, adult hemoglobin (Hb A), and fetal hemoglobin (Hb F) to produce 1 unspiked and 5 spiked samples per pool. The 35 laboratories participating in the survey used a total of 14 method-analyzer combinations. Measurements were performed in triplicate in a single run in accordance with manufacturer instructions. Absolute differences in creatinine concentration between spiked and unspiked samples were calculated per laboratory. Mixed ANOVA was used to quantify the interferent-related CV component for the Jaffe and enzymatic methods. RESULTS: The interference by bilirubin and Hb A on serum creatinine measurements was < 10% for most of the Jaffe and enzymatic methods. Obvious interference was observed among the Jaffe methods in samples spiked with Hb F, albumin, and IgG, but not among the enzymatic methods. The within-laboratory interferent-related CVs for the Jaffe method-analyzer combinations ranged from 8.0%-27% at a creatinine concentration of 40.4 mu mol/L (0.46 mg/dL) and from 5.4%-15% at 73.4 mu mol/L (0.83 mg/dL). Enzymatic methods had within-laboratory interferent-related CVs of < 4% at both concentrations. CONCLUSIONS: Albumin, IgG, and Hb F interfered with Jaffe creatinine assays, leading to inaccuracies in estimated glomerular filtration rates that are clinically important, especially in children and neonates. Because protein error and Hb F interference do not occur with any of the enzymatic methods tested, we conclude that enzymatic creatinine methods are preferred for evaluation of kidney function in pediatric cases. (c) 2008 American Association for Clinical Chemistry

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据