Standardization of C-peptide measurements

期刊

CLINICAL CHEMISTRY
卷 54, 期 6, 页码 1023-1026

出版社

AMER ASSOC CLINICAL CHEMISTRY
DOI: 10.1373/clinchem.2007.101287

关键词

-

资金

  1. PHS HHS [200200409985] Funding Source: Medline

向作者/读者索取更多资源

BACKGROUND: C-peptide is a marker of insulin secretion in diabetic patients. We assessed within- and between-laboratory imprecision of C-peptide assays and determined whether serum calibrators with values assigned by mass spectrometry could be used to harmonize C-peptide results. METHODS: We sent 40 different serum samples to 15 laboratories, which used 9 different routine C-peptide assay methods. We also sent matched plasma samples to another laboratory for C-peptide analysis with a reference mass spectrometry method. Each laboratory analyzed 8 of these samples in duplicate on each of 4 days to evaluate within- and between-day imprecision. The same 8 samples were also used to normalize the results for the remaining samples to the mass spectrometry reference method. RESULTS: Within- and between-run CVs ranged from < 2% to > 10% and from < 2% to > 18%, respectively. Normalizing the results with serum samples significantly improved the comparability among laboratories and methods. After normalization, the differences among laboratories in mean response were no longer statistically significant (P = 0.24), with least-squares means of 0.93-1.02. CONCLUSIONS: C-peptide results generated by different methods and laboratories do not always agree, especially at higher C-peptide concentrations. Within-laboratory imprecision also varied, with some methods giving much more consistent results than others. These data show that calibrating C-peptide measurement to a reference method can increase comparability between laboratories. (c) 2008 American Association for Clinical Chemistry.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据