4.3 Article

Wear behaviour in total ankle replacement: A comparison between an in vitro simulation and retrieved prostheses

期刊

CLINICAL BIOMECHANICS
卷 24, 期 8, 页码 661-669

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.clinbiomech.2009.06.006

关键词

Talar component; Three compartment ankle prosthesis; CMM; SEM; Raman spectroscopy

向作者/读者索取更多资源

Background: To minimise wear of the meniscal component in total ankle replacement, a three-component artificial joint has recently been developed. This new prosthesis has convex spherical tibial and anticlastic talar metal components with non-anatomic but ligament-compatible shapes in the sagittal plane, and a fully conforming ultra-high-molecular-weight-polyethylene meniscal component inserted in between. The in vitro wear of meniscal components can be assessed using a four-station joint simulator. The study was aimed at comparing wear patterns obtained in vitro with those observed in implant retrievals with the same design. Methods: The wear tests were run in a joint wear simulator at a frequency of 1.1 Hz for two million cycles. Three bearings within corresponding metal components were subjected to flexion/extension (range 0-58 degrees), anterior-posterior translation (0-5.2 mm), internal-external rotation (-1.9 degrees to +5.7 degrees), and a maximum axial load of 2.6 KN. These conditions were taken from the most recent findings in ankle joint mechanics. Three prostheses of the same type were harvested from patients due to replacement failures not associated with the device, 24, 24 and 9 months, respectively, after implantation. The in vitro worn components and the three retrievals were analysed by using a scanning electron microscope, a Coordinate Measuring Machine, and micro-Raman spectroscopy. Findings: Visual and microscopic observations, analyses, and Raman crystallinity-based measurements showed similarity between the patterns generated experimentally in the wear simulator and those seen in retrievals with similar wear life. Interpretation: A joint wear simulator like the one used in this study, once configured properly, appears to be suitable to assess wear rates also in total ankle prostheses. (C) 2009 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据