4.5 Article

Sample preparation for liquid chromatography-tandem mass spectrometry using functionalized ferromagnetic micro-particles

期刊

CLINICAL BIOCHEMISTRY
卷 41, 期 16-17, 页码 1417-1419

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2008.08.001

关键词

Chromatography; Extraction; Ferromagnetic micro-particles; Mass spectrometry; Sample preparation

向作者/读者索取更多资源

Objectives: To study the applicability of functionalized ferromagnetic micro-particles in the sample preparation for quantification of small molecules by LC-MS/MS. Design and methods: A representative analyte (itraconazole) was extracted from plasma samples using small quantities of C18-modified ferromagnetic micro-particles. Results: 125 mu g of functionalized micro-particles was sufficient to extract the target analyte from 10 mu L of plasma with a recovery rate of approximately 100%; linearity (r>0.99) and reproducibility (inter-assay CV <= 7%) of quantitative results was found. Conclusions: C18-functionalized ferromagnetic particles can be used to efficiently extract small molecule analytes from complex biological matrices for quantitative analyses using LC-MS/MS. (C) 2008 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.