4.6 Article

Self-Monitoring of Ambulatory Blood Pressure by the Microlife WatchBP O3-An Application Test

期刊

CLINICAL AND EXPERIMENTAL HYPERTENSION
卷 33, 期 1, 页码 34-40

出版社

TAYLOR & FRANCIS INC
DOI: 10.3109/10641963.2010.503300

关键词

blood pressure measurement; ambulatory blood pressure monitoring (ABPM); application test; Microlife WatchBP O3; hypertension

资金

  1. Ministry of Education, Culture, Sports, Science, and Technology, Japan [15790293, 16590433, 17790381, 18390192, 18590587, 19590929, 19790423]
  2. Ministry of Health, Labor and Welfare, Health and Labor Sciences Research Grants, Japan [H17-Kenkou-007, H18-Junkankitou [Seishuu]-Ippan-012, H20-Junkankitou [Seishuu]-Ippan-009, H20-Junkankitou [Seishuu]-Ippan-013]
  3. Japan Society for the Promotion of Science (JSPS) [16.54041, 18.54042, 19.7152, 20.7198, 20.7477, 20.54043]
  4. Ministry of Health, Labor and Welfare, Japan
  5. Japan Atherosclerosis Prevention Fund
  6. Uehara Memorial Foundation
  7. Takeda Medical Research Foundation
  8. National Cardiovascular Research Grants
  9. Biomedical Innovation Grants

向作者/读者索取更多资源

Ambulatory blood pressure monitoring (ABPM) is not widely used in clinical practice, because the measuring procedure is complex and the devices are expensive and need to be fitted by skilled medical technologists. The Microlife WatchBP O3 (Microlife AG, Widnau, Switzerland), which was developed for self blood pressure monitoring at home and adapted for ABPM, is highly affordable for an ABPM device and easy to manipulate. We performed an application test of the WatchBP O3 to confirm reliability of the device. Thirty-seven volunteer participants (age 30.4 +/- 13.5 y) underwent blood pressure (BP) measurements every 30 min for 24 h, and were asked to complete a questionnaire about the user-friendliness and acceptability of the device. The participants were asked to attach the device and to detach it the next morning by themselves. The quality of recordings was assessed in terms of percentage of valid readings. The mean number of 24-h BP readings per participant was 46.6 +/- 5.3. The percentage of valid readings was 90%, which was similar to that reported for traditional ambulatory devices. Eighty-six percent of participants found it is easy or very easy to attach the device by themselves. The WatchBP O3, which is easy to manipulate, may be convenient and acceptable for users. The percentage of valid readings is similar to that reported previously. A new era of self-monitoring of ambulatory BP is anticipated in the near future.