Evaluation of automated immunoassays for 25(OH)-vitamin D determination in different critical populations before and after standardization of the assays

期刊

CLINICA CHIMICA ACTA
卷 431, 期 -, 页码 60-65

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.cca.2014.01.026

关键词

Vitamin D; Standardization; Vitamin D-binding protein; Uremia

资金

  1. Abbott
  2. Roche

向作者/读者索取更多资源

Introduction: Standardization of immunoassays for 25(OH)-vitamin D determination is a major problem in clinical practice. A worldwide standardization program has started to address this and will reduce the bias observed between immunoassays. We aimed to calibrate 5 immunoassays on a LC-MS/MS traceable to the SRM 2972 and the ID-LC-MS/MS 25(OH)D Reference Method Procedure to see if the re-standardization would be efficient in a population of 3rd trimester pregnant women (PW), hemodialysis (HD) and osteoporosis (OP) patient. Material and methods: 184 serum samples (25(OH)D: 8.4-87 ng/ml) were selected to calibrate the immunoassays (Abbott-Architect, Roche-Elecsys, DiaSorin-Liaison, Siemens-Centaur and IDS-iSYS). Chromsystems MassChrom method was used as the referenced. Serum obtained in 34 PW, 25 HD and 34 OP patients were used as comparatives. Results: After adjusting to LC-MS/MS, immunoassays had regression slopes nearly identical to 1.0 with intercepts <0.5 ng/ml. However, in special populations, a systematic bias was still observed, except for iSYS. Conclusions: Re-standardization of 25(OH)D immunoassay will globally improve the differences. However, patients with a different serum matrix will still present significantly different results when they will be run with different methods. For those patients, the LC-MS/MS method seems to be the method of choice, even if some immunoassays are less influenced than others. (C) 2014 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据