Performance characteristics of a new Immulite® 2000 system erythropoietin assay

期刊

CLINICA CHIMICA ACTA
卷 412, 期 5-6, 页码 480-482

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.cca.2010.11.023

关键词

Imprecision; Method comparison; Interference; Chemiluminescent immunoassay; Recombinant EPO; Serum EPO; Immulite 2000

资金

  1. Siemens Healthcare Diagnostics
  2. ARUP Institute for Clinical and Experimental PathologyRegistered

向作者/读者索取更多资源

Background: Serum erythropoietin (EPO) measurements are useful for diagnostic evaluations of anemia, polycythemia, and other erythroid disorders. Methods: We evaluated a new formulation of the chemiluminescent immunoassay for EPO on the Immulite 2000 analyzer (Siemens Healthcare Diagnostics) for limit of blank (LoB), imprecision, linearity, interference, comparison to another commercially available assay, reference interval, and cross-reactivity with 2 recombinant EPO preparations. Results: The LoB was 0.23 IU/l. Total imprecision ranged from 4.2 to 12.1%. The assay was linear from 0 to 678 IU/l. Hemoglobin caused negative interference at concentrations >17.4 g/l. Deming regression from the method comparison study gave a slope of 1.00 +/- 0.04, an intercept of 3.9 +/- 12.7, and a S-y/x of 44.9 (r=0.98). The non-parametric reference interval was 33 to 23.4 IU/l. Epoetin alfa at a 1:2000 dilution gave mean results of 541 IU/l and 650 IU/l for Immulite 2000 and Access 2, respectively. Darbopoetin alfa at a 1:2000 dilution gave mean results of 337 IU/l and 579 IU/l for the Immulite 2000 and Access 2, respectively, indicating the 2 assays have substantially different cross-reactivities with recombinant EPO preparations. Conclusions: The new Immulite 2000 EPO assay shows acceptable performance and is suitable for routine clinical use. (C) 2010 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据