4.4 Article Proceedings Paper

Accurate calibration of test mass displacement in the LIGO interferometers

期刊

CLASSICAL AND QUANTUM GRAVITY
卷 27, 期 8, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0264-9381/27/8/084024

关键词

-

资金

  1. Direct For Mathematical & Physical Scien
  2. Division Of Physics [0757058] Funding Source: National Science Foundation

向作者/读者索取更多资源

We describe three fundamentally different methods we have applied to calibrate the test mass displacement actuators to search for systematic errors in the calibration of the LIGO gravitational-wave detectors. The actuation frequencies tested range from 90 Hz to 1 kHz and the actuation amplitudes range from 10(-6) m to 10(-18) m. For each of the four test mass actuators measured, the weighted mean coefficient over all frequencies for each technique deviates from the average actuation coefficient for all three techniques by less than 4%. This result indicates that systematic errors in the calibration of the responses of the LIGO detectors to differential length variations are within the stated uncertainties.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据