4.8 Article

Rates of and Factors Associated With Infection in 200 909 Medicare Implantable Cardioverter-Defibrillator Implants Results From the National Cardiovascular Data Registry

期刊

CIRCULATION
卷 130, 期 13, 页码 1037-1043

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1161/CIRCULATIONAHA.114.009081

关键词

implantable cardioverter-defibrillators; infection; infective endocarditis; registries; risk factors

资金

  1. American Heart Association
  2. American College of Cardiology Foundation's National Cardiovascular Data Registry (NCDR)

向作者/读者索取更多资源

Background-The rate of implantable cardioverter-defibrillator (ICD) infections has been increasing faster than that of implantation. We sought to determine the rate and predictors of ICD infection in a large cohort of Medicare patients. Methods and Results-Cases submitted to the ICD Registry from 2006 to 2009 were matched to Medicare fee-for-service claims data using indirect patient identifiers. ICD infections occurring within 6 months of hospital discharge after implantation were identified by ICD-9 codes. Logistic regression was used to examine factors associated with risk of ICD infection. Of 200 909 implants, 3390 patients (1.7%) developed an ICD infection. The infection rate was 1.4%, 1.5%, and 2.0% for single, dual, and biventricular ICDs, respectively (P<0.001). Generator replacement had a higher rate compared with initial implant (1.9% versus 1.6%, P<0.001). The factors associated with infection were adverse event during implant requiring reintervention (odds ratio [OR], 2.692; 95% confidence interval [CI], 2.304-3.145), previous valvular surgery (OR, 1.525; 95% CI, 1.375-1.692), reimplantation for device upgrade, malfunction, or manufacturer advisory (OR, 1.354; 95% CI, 1.196-1.533), renal failure on dialysis (OR, 1.342; 95% CI, 1.123-1.604), chronic lung disease (OR, 1.215; 95% CI, 1.125-1.312), cerebrovascular disease (OR, 1.172; 95% CI, 1.076-1.276), and warfarin (OR, 1.155; 95% CI, 1.060-1.257). Conclusions-Patients who developed an ICD infection were more likely to have had peri-ICD implant complications requiring early reintervention, previous valve surgery, device replacement for reasons other than battery depletion, and increased comorbidity burden. Efforts should be made to carefully consider when to reenter the pocket at any time other than battery replacement.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据