期刊
CHEST
卷 137, 期 5, 页码 1108-1115出版社
ELSEVIER
DOI: 10.1378/chest.09-1504
关键词
-
资金
- National Heart, Lung, and Blood Institute [HL-52586]
Background: Theoretical considerations and limited scientific evidence suggest that whole-body plethysmography overestimates lung volume in patients with severe airflow obstruction. We sought to compare plethysmography (Pleth)-, helium dilution (He)- and CT scan-derived lung volume measurements in a sample containing many patients with severe airflow obstruction. Methods: We measured total lung capacity (TLC) in 132 patients at three hospitals, with monitored application of recommended techniques for Pleth and He measurements of lung volume and by thoracic CT scans obtained during breath hold at full inspiration. Results: Average TLC among 132 subjects was 6.18 L (+/- 1.69 L) by Pleth-derived TLC, 5.55 L (+/- 1.39 L) by He-derived TLC, and 5.31 L (+/- 1.47) by CT scan-derived TLC. Pleth-derived TLC was significantly greater than either He-derived TLC or CT scan-derived TLC (P <= .001), whereas there was no significant difference between He-derived and CT scan-derived values. When examined separately, there were significant within-subject differences in TLC by measurement technique among subjects with airflow obstruction, but not among those without airflow obstruction. Plethysmographic overestimation of TLC was greatest among subjects with FEV1 <30% of predicted. Conclusions: In the setting of airflow obstruction, Pleth systematically overestimates lung volume relative to He or thoracic imaging despite adherence to current recommendations for proper measurement technique. CHEST 2010; 137(5):1108-1115
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐