期刊
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
卷 105, 期 5, 页码 689-700出版社
US GOVERNMENT PRINTING OFFICE
DOI: 10.6028/jres.105.055
关键词
cryogenic radiometer; detector standards; electrical substitution; IR detector; near infrared; radiometry; responsivity
A cryogenic radiometer-based system was constructed at the National Institute of Standards and Technology for absolute radiometric measurements to improve detector spectral power responsivity scales in the wavelength range from 900 nm to 1800 nm. In addition to the liquid-helium-cooled cryogenic radiometer, the system consists of a 100 W quartz-tungsten-halogen lamp light source and a 1 m single-grating monochromator for wavelength selection. The system was characterized and the uncertainty in spectral power responsivity measurements evaluated. A variety of photodetectors, including indium gallium arsenide photodiodes (InGaAs), germanium (Ge) photodiodes, and pyroelectric detectors, were subsequently calibrated. Over most of the spectral range, the spectral power responsivity of the photodetectors can be measured with a combined relative standard uncertainty of 0.4 % or less. This is more than a factor of two smaller than our previous capabilities, and represents a significant improvement in the near infrared (NIR) spectral power responsivity scale maintained at NIST. We discuss the characterization of the monochromator-based system and present results of photodetector spectral power responsivity calibrations.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据