JOURNAL OF MICROELECTROMECHANICAL SYSTEMS
卷 9, 期 1, 页码 9-17出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/84.825771
关键词
force transducer; heart cell; hinged microstructures; MEMS; micromechanical; surface micromachining
A microelectromechanical systems (MEMS) force transducer system, with a volume less than 1 mm(3) millimeter, has been developed to measure forces generated by living heart muscle cells. Cell attachment and measurement of contractile forces have been demonstrated with a commercially fabricated surface-micromachined hinged polysilicon device, Two freestanding polysilicon clamps, each suspended by a pair of microbeams, hold each end of a heart cell. When the cell contracts, the beams bend and force is determined from the measured deflection and the spring constant in the beams. The average maximal force over seven contractile experiments using a calcium solution stimulus was F-max = 12.6 +/- 4.66 mu N. Normalizing to a cross-sectional area, F-max/area was 23.7 +/- 8.6 mN/mm(2). These force data were also correlated to optically imaged striation pattern periodicity. Intermediate forces were also measured in response to a calcium solution gradient and showed similar behavior to those measured in other laboratories. This MEMS force transducer demonstrates the feasibility of higher fidelity measurements from muscle cells and, thus, an improved understanding of the mechanisms of muscle contraction. [389].